Appendix 1, Technical Data
186
MC4 User Guide
Frequency Measurement
Full Range: 0.0027 50 000 Hz
Sub Range
Resolution
1 Year Uncertainty
(1
0.0027 .. 0.5 Hz
0.000001 Hz
0.01 % RDG
0.5 .. 5 Hz
0.00001 Hz
(same as above)
5 .. 50 Hz
0.0001 Hz
(same as above)
50 .. 500 Hz
0.001 Hz
(same as above)
500 .. 5000 Hz
0.01 Hz
(same as above)
5000 .. 50000 Hz
0.1 Hz
(same as above)
Feature
Specification
Temperature
coefficient
Specification valid from –10 6 50°C
(14 6 122°F)
Input impedance
> 1 Mohm
Trigger level
-1614 V in 1 V steps and open collector inputs
Minimum signal
amplitude
2 Vpp (< 10 kHz),
3 Vpp (10...50 kHz)
Supported units
Hz, kHz, cph, cpm, 1/Hz (s), 1/kHz (ms), 1/MHz (µs)
Gate period
267 ms + 1 signal period
Pulse Counting
Feature
Specification
Range
0 to 9,999,999 pulses
Input impedance
> 1 Mohm
Trigger level
-1614 V in 1 V steps and open collector inputs
Minimum signal
amplitude
2 Vpp (pulse length > 50 µs)
3 Vpp (pulse length 10...50 µs)
Switch Test
Potential free contacts
Test Voltage
(Trigger level)
3 V 0.13 mA (1 V) or
24 V 35 mA (2 V)
Voltage level detection
Input impedance
> 1 Mohm
Trigger level
-1614 V in 1 V steps
1)
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity,
repeatability and typical long term stability for mentioned period. (k=2)
Summary of Contents for MC4
Page 46: ...Service 34 MC4 User Guide Empty ...
Page 106: ...Tools Menu 94 MC4 User Guide Empty ...
Page 134: ...Related Information 122 MC4 User Guide Empty ...
Page 195: ...Appendixes Appendix 1 Technical Data 184 Appendix 3 Index 203 ...
Page 223: ...Notes MC4 User Guide 211 Notes ...
Page 224: ...Notes 212 MC4 User Guide ...