Editing Instrument Data
MC4 User Guide
169
FIELD
(cont.)
DESCRIPTION
Final Ramp Delay
[ms]
This setting defines the ramp's change rate
during the final search for set and reset
point.
Try, e.g. using a 30 millisecond delay . If
the detected set and reset points seem too
inconsistent, increase this value and try
again. Keep in mind, however, that a de-
fective switch is always inconsistent, no
matter how you increase the delay.
Switch Test Cycles
How many test cycles (set/reset) is done
for each As Found or As Left calibration
run.