Chapter 6: Recording and analyzing data
89
Preparing the workspace
Introduction
This topic describes how to prepare the workspace and apply
application settings to your experiment before recording data.
Procedure
To prepare the workspace:
1. Using the
Browser
toolbar, create a new specimen in
MyExperiment
and rename it
FourColorBeads
.
2. Create two tubes for the
FourColorBeads
specimen. Rename
the tubes
Beads_001
and
Beads_002
.
3. Expand the
Global Worksheets
folder in
MyExperiment
to
access the default global worksheet, and rename the worksheet
MyData
.
4. On the
MyData
worksheet, create the following plots for
previewing the data:
• FSC vs SSC
• FITC vs PE
• FITC vs PerCP-Cy5.5
• FITC vs APC
Applying saved
application settings
to a new
experiment
When applications settings are applied to an experiment, the
cytometer settings are updated with the parameters included in the
application settings, optimized PMT voltages, threshold settings,
area scaling factors, and window extension values.
To apply saved application settings to your experiment:
1. Right-click the experiment-level Cytometer Settings and select
Application Settings > Apply.
Summary of Contents for FACSymphony A3
Page 4: ......
Page 8: ...User s Guide 8...
Page 14: ...This page intentionally left blank...
Page 46: ...This page intentionally left blank...
Page 86: ...This page intentionally left blank...
Page 97: ...Chapter 6 Recording and analyzing data 97 e Click OK 12 Print the analysis...
Page 98: ...User s Guide 98 Your global worksheet analysis objects should look like the following...