User’s Guide
118
8. Install the FITC-positive control tube onto the loading port
and click
Load
in the
Acquisition Dashboard.
9. Adjust the FSC and SSC voltages to place the particles on
scale.
10. Adjust the P1 gate around the population of interest.
11. Adjust the FSC area scaling.
a. Click the
Laser Tab
in the
Cytometer
window.
b. Adjust the FSC area scaling factor until the FSC-A signal
matches the FSC-H signal:
• Increase the area scaling factor if the FSC-A signal is
lower than FSC-H.