Changing the Bal seal
Introduction
This topic describes how to replace the Bal seal.
The sample injection tube Bal seal is a ring that forms a seal with
the sample tube and ensures proper tube pressurization.
When to change
the Bal seal
Over time, the Bal seal becomes worn or cracked and requires
replacement. Replacement is necessary if a proper seal is not
formed when a sample tube is installed on the SIP. Indications that
a proper seal has not formed include:
•
The tube will not stay on the SIP without the tube support
arm.
•
When the tube is installed and RUN is pressed on the
cytometer, the RUN button is orange (not green).
Procedure
To replace the Bal seal:
1. Remove the outer sleeve from the sample injection tube by
turning the retainer counterclockwise. Slide the outer sleeve
down and off of the sample injection tube.
Retainer
Outer sleeve
Sample injection tube
Bal seal
Work carefully. The outer sleeve can fall off as you loosen the
retainer.