![BD FACSymphony A3 User Manual Download Page 113](http://html1.mh-extra.com/html/bd/facsymphony-a3/facsymphony-a3_user-manual_2714976113.webp)
Chapter 8: Manual settings
113
Optimizing laser delay
Introduction
This topic describes how to optimize the laser delay using
BD FACSDiva software.
Before you begin
To optimize the delay for a given laser, you must acquire events
from a sample with a fluorescence signal excited by that laser.
Follow the procedures in
Recording and analyzing data (page 87)
,
for sample optimization and acquiring data.
Procedure
To optimize laser delay:
1. While acquiring data from your sample, create a histogram to
show the fluorescence signal excited by the laser for which the
delay is to be optimized.
2. In the
Acquisition Dashboard
, set the
Events to Display
to
500
events.
3. Click the
Laser
tab in the
Cytometer
window.
Window extension and laser delay values are displayed in
microseconds (µs).
4. Set the
Window Extension
value to
0
µs.