Chapter 7: Troubleshooting
105
No events in
acquisition display
and RUN button is
orange
No fluorescence
signal
Possible causes
Recommended solutions
RUN is not activated
Press the RUN button.
Sample tube is not installed
or is not properly seated
Install the sample tube correctly on the
SIP.
Sample tube is cracked
Replace the sample tube.
Waste tubing line is
disconnected from the waste
container tank
Connect the waste tubing line to the
waste container tank.
Sheath container is not
pressurized
Ensure that the sheath container lid
and all connectors are securely
seated.
Inspect the sheath container O-ring
inside the lid and replace it if
necessary. See
replacing the sheath gasket
(page 62)
Bal seal is worn
Replace the Bal seal. See
Air leak at sheath container
Ensure that the sheath container lid
and all connectors are securely seated.
Sheath container is empty
Fill the sheath container.
Air in sheath filter
Purge the filter. See
.
Possible causes
Recommended solutions
Incorrect fluorochrome
assignment
Make sure that the cytometer
configuration in the software matches
the optical filters in the cytometer and
the configuration is as expected.
Laser is not functioning
Call your BD service representative.
Summary of Contents for FACSymphony A3
Page 4: ......
Page 8: ...User s Guide 8...
Page 14: ...This page intentionally left blank...
Page 46: ...This page intentionally left blank...
Page 86: ...This page intentionally left blank...
Page 97: ...Chapter 6 Recording and analyzing data 97 e Click OK 12 Print the analysis...
Page 98: ...User s Guide 98 Your global worksheet analysis objects should look like the following...