Chapter 5: Optimizing cytometer settings
71
If any parameters did not pass, see the
BD Cytometer Setup
and Tracking Application Guide
for troubleshooting
information.
12. Select
File > Exit
to close the CS&T window and return to the
BD FACSDiva interface.
The CST Mismatch dialog opens.
Click the
Details
button to verify which cytometer settings will
be updated.
13. Click
Use CST Settings
.
By selecting
Use CST Settings
, the laser delay, area scaling, and
other cytometer settings will be updated to the latest settings
from the performance check.
Next step
Continue the optimization of your cytometer for an experiment or
sample type as described in
Setting up an experiment (page 72)
.