Chapter 3: Cytometer setup
39
3. Direct the vent line into a beaker and press the small button at
the end of the vent fitting against the side of the beaker until a
steady stream of fluid empties from the filter.
Button
Vent fitting
4. Tilt the filter and verify that no trapped air remains in the
filter.
5. Repeat steps
and
until no air is observed in the filter.
6. Check the sheath line for air bubbles.
7. Open the roller clamp at the fluidics interconnect (if necessary)
to bleed off any air in the line. Collect any excess fluid in a
waste container.
8. Close the roller clamp.