ULTRASAMPLER® sample collection system
192
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
26.10.2
Changing the pump hose
The pump hose can be changed at any time but not while the sampler is in
operation.
Removing the old pump hose
Undo star handle and remove rotor protection (Plexiglas).
Unscrew union nut on injection needle and intake cannula.
Take out old pump hose, remove union nut.
Fitting the new pump hose
NOTE:
Make sure that the new pump hose is not twisted when it is fitted.
Check that the sample bottles are being filled properly after changing the
hose.
Fit new pump hose on injection needle and tighten union nut.
Put the conveying wheel in the position shown in the diagram and fit
pump hose in the groove opposite.