Start of operation
203
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
Start delay setting in case of below-average fat values:
The start of the sampler can be changed with this parameter. The milk at
the start of the intake is low in fat therefore, by delaying the start of the
sampler, less low-fat milk is taken into the bottle, leading to a general in-
crease in the fat content in the sample.
Start delay parameter in MAK 3003.
Configuration - Periphery - ULTRASAMPLER:
Start delay: 0.4 s
(Start with a value between 0.4 and 0.7, run a test with one quantity of milk,
take the fat reading, adjust the setting if necessary and run the test again!)
In case of above-average fat values:
Option 1 (only in exceptional cases and with milk collection vehicles with
high throughputs):
Above-average fat levels were measured in the higher quantities (80 l and
160 l) in the case of milk collection vehicles with high throughputs in con-
junction with the ULTRASAMPLER system therefore, although there was a
slightly lower than average fat level in the small quantity (40 l), the following
changes were introduced (from software version 2.32 of the
ULTRASAMPLER controller).
3 new parameters were introduced (1 per intake quantity), allowing the level
curve to be set between proportional (more fat) and 1/X (less fat) on the
ULTRASAMPLER controller.
The 3 parameters in sample collection/basic settings are defined as follows:
Low correction factor:
Effective for intake quantities which are below and equal to the low/medium
switching threshold.
Medium correction factor:
Effective for intake quantities which are between the low/medium and me-
dium/high switching thresholds.
High correction factor:
Effective for intake quantities which are above and equal to the medium/high
switching threshold.