ULTRASAMPLER® sample collection system
193
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
Continue turning the conveying wheel clockwise and fitting the pump
hose until the pump hose is fully fitted in the groove.
Place union nut on pump hose.
Fit pump hose on intake cannula and tighten union nut.
Fit rotor protection and screw down star handle.