Figure 12 - Finger Attachment-Screw Location
4.
Gently tilt the finger slightly forward and lift the alignment “teeth” out of their slots as shown
in Figure 13. If the joint-torque sensor option is installed, BE CAREFUL not to damage the
gold-plated electrical contact pins when disengaging the teeth.
5.
Once the teeth are disengaged, move the finger away from its motor along a straight line
perpendicular to the motor’s face. Do not twist or rock the finger when removing or attaching
it.
Open a Small
Gap (0.5mm
Max)
Lift
Alignment
Teeth out of
Slots
Pull Finger Away
from Motor Face
Figure 13 - Removing the Fingers for Maintenance
6.
Make note of the relative position of the fingertip and inner finger link since removing the
finger will disengage the coupling between them. Should either link, or the spur gears to
which they are attached, move after the finger has been removed, the fingertip position must
be reset. Use a 2-mm hex wrench to manually rotate the Joint-3 drive 5 1/2 revolutions from
the position where both links are inline and horizontal.