17. Troubleshooting
Code E4000000: “Pump 1 speed too low” (Error)
Situation
Solution
Pump 1 (DMD cooling circuit) not
running at expected speed.
Check whether the pump is properly connected towards the Thermal hub board.
Damaged pump
Replace the DMD cooling assembly.
Code E4000100: “Pump 2 speed too low” (Error)
Situation
Solution
Pump 2 (Light Source) not running
at expected speed.
Check whether the pump is properly connected towards the Thermal hub board.
Code E4000200: “Mainboard FPGA Fan speed too low” (Error)
Situation
Solution
Blocked fan
Unblock fan fpga on Support frame of the Mainboard. Ensure that the fan can turn freely.
Wiring issue
Remove the mainboard from the projector. Check / reseat the wire unit of the fan. If
problem persists, replace the fan.
Malfunction Mainboard
Replace the Main Board.
Fan end of life.
Replace the fan.
Code E4000300: “ Mainboard PCB Fan speed too low” (Error)
Situation
Solution
Blocked fan
Unblock fan pcb on Support frame of the Mainboard. Ensure that the fan can turn freely.
Wiring issue
Remove the mainboard from the projector. Check / reseat the wire unit of the fan. If
problem persists, replace the fan.
Malfunction Mainboard
Replace the Main Board. .
Fan end of life.
Replace the fan. .
Code E4000400: “Radiator Fan 1 speed too low” (Error)
Situation
Solution
Blocked fan
Remove the fan from the radiator assembly and verify if the fan is not blocked. To access
the fan, see service procedure
Wiring issue
Check the connection towards the thermal hub board. If problem persists, replace the
fan.
Fan end of life.
Replace the fan. .
Code E4000500: “Radiator Fan 2 speed too low” (Error)
Situation
Solution
Blocked fan
Remove the fan from the radiator assembly and verify if the fan is not blocked. To access
the fan, see service procedure .
Wiring issue
Check the connection towards the thermal hub board. If problem persists, replace the
fan.
Fan end of life.
Replace the fan. See service chapter
Code E4000600: “Radiator Fan 3 speed too low” (Error)
Situation
Solution
Blocked fan
Remove the fan from the radiator assembly and verify if the fan is not blocked.
Wiring issue
Check the connection towards the thermal hub board. If problem persists, replace the
fan.
Fan end of life.
Replace the fan.
72
723–0016 F90 01/12/2017
Summary of Contents for F90 series
Page 1: ...F90 Service Manual 723 0016 02 01 12 2017...
Page 6: ...Table of contents 4 723 0016 F90 01 12 2017...
Page 8: ...1 Introduction 6 723 0016 F90 01 12 2017...
Page 10: ...2 General safety 8 723 0016 F90 01 12 2017...
Page 18: ...4 DLP System General Description 16 723 0016 F90 01 12 2017...
Page 25: ...5 GP6 system functional description Image 5 5 723 0016 F90 01 12 2017 23...
Page 28: ...5 GP6 system functional description 26 723 0016 F90 01 12 2017...
Page 34: ...7 OSD display menu 32 723 0016 F90 01 12 2017...
Page 36: ...8 Thermal management system Image 8 1 34 723 0016 F90 01 12 2017...
Page 42: ...8 Thermal management system 40 723 0016 F90 01 12 2017...
Page 58: ...14 XPS actuator 56 723 0016 F90 01 12 2017...
Page 60: ...15 Light engine 58 723 0016 F90 01 12 2017...
Page 82: ...18 Assembly Hierarchy 80 723 0016 F90 01 12 2017...
Page 103: ...20 Scheduled operations Image 20 42 Image 20 43 723 0016 F90 01 12 2017 101...
Page 104: ...20 Scheduled operations 102 723 0016 F90 01 12 2017...
Page 112: ...21 DMD DMD Board CLGA DMD heatsink elements 110 723 0016 F90 01 12 2017...
Page 132: ...24 Lens shift assembly 130 723 0016 F90 01 12 2017...
Page 134: ...25 Prism Housing Image 25 5 132 723 0016 F90 01 12 2017...