64
Coronis 5MP
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the B
ARCO
C
ORONIS
Immunity
test
IEC 60601
test level
Level
Electromagnetic environment –
guidance
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Not
applic
able
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V
3 V/
m
Portable and mobile RF communications
equipment should be used no closer to any
part of the B
ARCO
C
ORONIS
Display, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with symbol:
P
2
.
1
d
P
2
.
1
d
P
3
.
2
d
Summary of Contents for Coronis 5MP Mammo
Page 1: ...User Guide Coronis 5MP...
Page 2: ...2 Coronis 5MP...
Page 4: ...4 Coronis 5MP...
Page 87: ...Coronis 5MP 87...