Coronis 5MP
45
T
O
SELECT
A
READING
ROOM
:
1
Bring up the OSD main menu.
2
Navigate to the
Configuration
>
Calibration > Ambient Light
menu.
3
Enter the
Reading Room
submenu.
4
Select one of the available reading rooms and confirm.
Continuous ALC
A
BOUT
CONTINUOUS
ALC
Enabling continuous ALC will continuously recalculate the DICOM display
function taking the averaged ambient light into account.
T
O
SELECT
CONTINUOUS
ALC
:
1
Bring up the OSD main menu.
2
Navigate to the
Configuration
>
Calibration > Ambient Light
menu.
3
Enter the
Continuous ALC
submenu.
4
Select
Enabled/Disabled
as desired and confirm.
Emergency Room
Corresponds to light conditions in
emergency rooms.
Operating Room
Corresponds to light conditions in operating
rooms. This setting has the highest
maximum ambient light.
Continuous ALC can only be selected when the DICOM display
function is selected. Therefore, please refer to
“Display
functions”
on page 42 to correctly set the display function.