Coronis 5MP
63
ELECTROMAGNETIC
IMMUNITY
The B
ARCO
C
ORONIS
Display is intended for use in the electromagnetic
environment specified below. The customer or the user of the B
ARCO
C
ORONIS
Display should assure that it is used in such an environment.
U
T
is the a.c. mains voltage prior to application of the test level
Immunity test
IEC 60601
Test levels
Compliance
level
Electromagnetic
environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply lines
± 1kV for input/
output lines
± 2kV for power
supply lines
± 1kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment
Surge
IEC61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality
should be that of a
typical commercial or
hospital environment
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95%
dip in U
T
) for 0.5
cycle
40% U
T
(60%
dip in U
T
) for 5
cycles
70% U
T
(30%
dip in U
T
) for 25
cycles
< 5% U
T
(>95%
dip in U
T
) for 5s
< 5% U
T
(> 95%
dip in U
T
) for 0.5
cycle
40% U
T
(60% dip
in U
T
) for 5 cycles
70% U
T
(30% dip
in U
T
) for 25
cycles
< 5% U
T
(>95%
dip in U
T
) for 5s
Mains power quality
should by that of a
typical commercial or
hospital environment. If
the user of the B
ARCO
C
ORONIS
Display
requires
continued operation
during power mains
interruptions, it is
recommended that the
B
ARCO
C
ORONIS
Display be
powered from an
uninterruptible power
supply or a battery.