Coronis 5MP
61
EMC notice
No specific requirement on the use of external cables or other
accessories except power supply.
With the installation of the device, use only the delivered power supply
or a spare part provided by the legal manufacturer. Using another can
result in a decrease of the immunity level of the device.
ELECTROMAGNETIC
EMISSIONS
The B
ARCO
C
ORONIS
Display is intended for use in the electromagnetic
environment specified below.
The customer or the user of the B
ARCO
C
ORONIS
Display should assure that
it is used in such an environment.
This B
ARCO
C
ORONIS
Display complies with appropriate medical EMC
standards on emissions to, and interference from surrounding
equipment. Operation is subject to the following two conditions: (1) this
device may not cause harmful interference, and (2) this device must
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The B
ARCO
C
ORONIS
Display uses RF energy
only for its internal function. Therefore, its
RF emissions are very low and are not likely
to cause any interference in nearby
electronic equipment
RF emissions
CISPR 11
Class B
The B
ARCO
C
ORONIS
Display
is suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies