Action
Result
Navigate: MENU > Dx_OUT > Tch2Pt > TchSPt1
OR
Navigate: MENU > Dx_OUT > Tch2Pt > TchSPt2
The selected TEACH mode and "Teaching"
display while the sensor is being taught.
TEACH Accepted
The new value is shown on the second line
of the display and flashes before it is saved
and the sensor returns to the parent menu.
TEACH Not Accepted
"FAIL" and a warning message display, and
the sensor returns to the parent menu.
3. Repeat steps 1 to 2 for the other switch point, if desired.
Remote Input Instructions
1. Access the TEACH mode.
Action
Result
Single-pulse the remote input.
T
"RMT TCH" and the current switch point
value displays.
2. Present the target.
Action
Result
Present the switch point one target.
"RMT TCH" and the target's measurement
value display.
3. TEACH the sensor.
Action
Result
Double-pulse the remote input.
T
T
T
"TchSPt1 Teaching" displays while the
sensor is being taught.
TEACH Accepted
The new value displays on the second line
of the display, flashes, and the sensor goes
to "TchSPt2" and the current
measurement value.
TEACH Not Accepted
"FAIL" flashes, the sensor returns to step
2, and "RMT TCH" displays.
4. Present the target.
Action
Result
Present the switch point two target.
"TchSPt2" and the target's measurement
value display.
5. TEACH the sensor.
Action
Result
Single-pulse the remote input.
T
"TchSPt2 Teaching" displays while the
sensor is being taught.
TEACH Accepted
The new value displays on the second line
of the display, flashes, and the sensor
returns to Run mode.
TEACH Not Accepted
"FAIL" flashes, the sensor returns to step
2, and "RMT TCH" displays.