4.3.2 Connection of pH cable 9060 to terminal of BAMOPHAR
1°) Remove the BAMOPHAR lower cover to access to the terminal block.
2°) Remove the lid of the terminal, sliding it downwards or upwards.
Coaxial cable: Core terminal
Clamp for copper shield
Lid
3°) Pass the prepared cable through the cable gland.
4°) Insert the tip in the screw connector
5°) Be sure that the electrical contact between the copper shield and the 2 brackets will be perfect.
6°) Press on the cable to insert it properly in the brackets.
7°) Screw on tightly to block the tip in its connector.
8°) Replace the lid of the terminal in its initial position.
10°) Fit back the lid of the BAMOPHAR.
9°) Tightly screw on the cable gland to assure a watertight connection.