SETTING OF THE RELAY REGULATION MODE

Go to the menu RELAY REGULATION

• There is no access to settings of relay regulation mode if relays S1 and/ or S2 are operating; So, go to::

ON / OFF MODE

To access to this mode, first go back to menus ADJUST ALARM S1 and S2 to set them in OFF
mode.

• There is no access to relay regulation mode when PID regulation is operating; So go to:

PID MODE

To access to relay regulation mode, go to to the menu REGUL. PID to switch it OFF.

• Regulation mode is thus accessible if, and only if, the THRESHOLD and PID modes are deactivated.

REGULATION ON / OFF

Select ON mode to set up parameters, then confirm.

SET VAL 07.00 pH

Enter the threshold value, then confirm.

T. CYCLE 0060 Sec

Enter the cycle time according to the process, then confirm.

HIGH PB 00.80 pH

Enter the differential for upper bandwidth, then confirm.

Note: Output S2 is dedicated for upper bandwidth.

LOW PB 00.50 pH

Enter the differential for lower bandwidth, then confirm.

Note: Output S1 is dedicated to lower bandwidth.

HIGH DZ 00.50 pH

Enter the differential value for the upper dead zone, then confirm.

LOW DZ 00.20 pH

Enter the differential value for the lower dead zone, then confirm.

SAVE ?

To record the settings, press the icon SAVE.

Cycle time

High limit

Set point

Low limit

Time

S1 actuated

No action

Dead zone

Measurement

S2

EXAMPLE:

For settings such as following:
- Set point: pH 7.00
- Cycle time: 10 s

- Upper bandwidth: 3 pH (i.e. high limit is pH 10.00)
- Lower bandwidth: 1 pH (i.e. lower limit is pH 6.00)
- High dead zone: 0.3 pH (from pH 7.00 to 7.30)
- Lower dead zone: 0.7 pH (from pH 6.30 to 7.00)

Results for operating regulation:

- Above the High limit, between pH 10 and 14, output S2 is continuously activated (continuous injection of base, pH will decrease).
- Below the Low limit, between pH 0 and 6, output S1 is continuously activated (continuous injection of acid, pH will increase).
- Within the Dead zones, between pH 6.3 and 7.3, the outputs S1 and S2 are not actuated.

- When measurement is above Dead zone and below High limit (pH 7.3 to 10), or below Dead zone and above the Low limit (pH 6 to 6.3), the

outputs S1 or S2, are actuated during a time proportional to the difference between set point and measurement.

Actuation time = Cycle time x (Measurement - Set point) / High or Low bandwidth

Caution:

According to the above example, if measurement is pH 7.80, actuation time = 3 s: 2.66 = 10 x (7.8 - 7) / 3

MES

23-11-2018

M-107.01-EN-AE

107-01

/13

pH/ mV monitors

BAMOPHAR 107

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Summary of Contents for BAMOPHAR 107

Page 1: ...MANUAL MES 23 11 2018 M 107 01 EN AE 107 01 1 pH mV monitors BAMOPHAR 107 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www bamo...

Page 2: ...RELAY OUTPUTS S1 S2 AND S3 AS THRESHOLDS 9 2 1 11 COMMENTS ABOUT REGULATION WITH BAMOPHAR 9 2 2 11 ADJUSTMENT OF THRESHOLDS S1 S2 AND S3 IN MODE ON OFF REGULATION 9 2 3 11 ASSIGNMENT OF S3 THRESHOLD...

Page 3: ...sensor or cleaning mode Contact Initial resistance 100 m max voltage drop 6 V DC 1 A Switching power 831 VA AC 3 A 277 V AC 90 W 3 A 30 V DC Switching capacity min 100 mA 5 V DC depending of frequency...

Page 4: ...top cover Measurement input Screw connector for pH cable type 9060 Note Signal input on screw connector for 9060 cable type Only for wall mount models codes 107520 107521 107523 and 107524 MES 23 11...

Page 5: ...O contact for measurement temperature 21 S1 22 Relay S3 N O contact for measurement or temperature 23 S3 May be assigned to the external signal becomes NC contact see 9 2 3 24 25 Main power supply 26...

Page 6: ...ayed as pH 7 4 3 1 9060 coaxial cable preparation 1 4 2 5 6 3 Remove the outer insulation by 20 mm Fold back the shield copper mesh Remove the 2nd sheath black layer over 20 mm Strip the core of the c...

Page 7: ...ontact between the copper shield and the 2 brackets will be perfect 6 Press on the cable to insert it properly in the brackets 7 Screw on tightly to block the tip in its connector 8 Replace the lid of...

Page 8: ...tatus of relays S1 S2 S3 S4 Designations Identifications of your instruments Measurements display Parameter settings of both BAMOPHAR Display settings Date and time Measurements details Opened or clos...

Page 9: ...Confirm setting and go to the next parameter SAVE To save the parameters Closed means CONSULTATION review mode e g Green icon for extension unit 7 DISPLAY SETTINGS Note To access to the display setti...

Page 10: ...buffer pH 4 00 for a process with acids Press MENU and choose ADJUST ELECTRODE STANDARD 7 00 pH Plunge the tip of the electrode in the pH 7 00 buffer solution Ref 9011 then confirm Do not forget to re...

Page 11: ...e the relay S1 select OFF mode then confirm and save the setting ALARM 1 MEASURE TEMP In ON mode threshold S1 can be set in MEASUREMENT or TEMPERATURE mode MEASURE mode threshold dedicated to pH or mV...

Page 12: ...UT 4 20 mA or 0 20 mA Select the right scale 4 20 mA or 0 20 mA for output signal then confirm SAVE To record the settings press the icon SAVE 9 5 SETTINGS OF ANALOGUE OUTPUT FOR TEMPERATURE Go to men...

Page 13: ...tings press the icon SAVE Cycle time High limit Set point Low limit Time S1 actuated No action Dead zone Measurement S2 EXAMPLE For settings such as following Set point pH 7 00 Cycle time 10 s Upper b...

Page 14: ...o go to ADJUST ALARM To access to the CLEANING PROBE menu switch alarms mode to WITHOUT ALARM in menu ADJUST ALARM If the ALARM mode is not activated the CLEANING PROBE function is available CLEANING...

Page 15: ...Ti s Td s PID 1 2 x T t 2 x t 0 5 x t PI 0 9 x T t 3 3 x t 00 P T t 9999 00 PID analogue output 8 4 In MANUAL mode REGUL MANU pH measurement Reached set point Time Time Cde 10 m 10 To To 9 10 TYPE OF...

Reviews: