4.3
CONNECTION OF pH CABLE 9060 TO WALL MOUNT BAMOPHAR
Only for wall mount models - codes 107520/ 21/ 22/ 23/ 24
Measuring signal is safely transmitted through a special low noise cable (Ref. 9060)
The insulation must be perfect between the core and the shield. Humidity traces are prohibited.
IMPORTANT: a defect on insulation or a short-circuit will be displayed as pH 7.
4.3.1 9060 coaxial cable preparation
1.
4.
2.
5.
6.
3.
Remove the outer insulation by 20 mm.
Fold back the shield (copper mesh)
Remove the 2nd sheath (black layer) over 20 mm.
Strip the core of the coaxial cable for 10 mm
Recommended: Crimp a wiring tip (1 mm², length 8 mm,
red color)
Shorten the tip (length of 4 mm is necessary).
MES
23-11-2018
M-107.01-EN-AE
107-01
/6
pH/ mV monitors
BAMOPHAR 107