pH RISING

pH DECREASING

Trigger point RELAY ON

Trigger point RELAY OFF

Time

pH

pH

Time

8.5

7

Relay actuated

Relay actuated

Delay before relay switches ON

Actuation of S2

pH

pH

Trigger point RELAY OFF

Trigger point RELAY ON

Time

Time

Relay ON

Relay ON

Delay before relay switches OFF

8.5

7

9.3

SETTING OF TEMPERATURE

Go to the menu TEMPERATURE

MEASURE: AUTO / MANUAL

AUTO : With signal from a sensor Pt 100 Ohm at 0°C

MANUAL : Without sensor Pt 100 Ω; temperature value is entered manually.
Confirm the selection.

FLUID T. + 025.0°C

If the MANUAL mode is selected, enter the temperature value of the fluid, in °C, then confirm.

SAVE ?

To record the settings, press the icon SAVE.

9.4

SETTINGS OF ANALOGUE OUTPUT FOR MEASUREMENT

Go to menu OUTPUT mA pH (or RH for ORP)

HIGHER 14.00 pH

Enter the highest value for pH (or ORP) for the output signal at 20.00 mA, then confirm.

LOWER 00.00 pH

Enter the lowest value for pH (or ORP) for the output signal at 4.00 mA (or 0.00 mA), then confirm.

OUTPUT 4/20 mA or 0/20 mA

Select the right scale 4...20 mA or 0...20 mA for output signal, then confirm.

SAVE ?

To record the settings, press the icon SAVE.

9.5

SETTINGS OF ANALOGUE OUTPUT FOR TEMPERATURE

Go to menu OUTPUT mA TEMP.

160.0°C

Enter the temperature value in °C for output signal value at 20.00 mA, then confirm.

LOWER +000.0°C

Enter the temperature value in °C for output signal value output signal at 4.00 mA (or 0.00 mA),
then confirm.

OUTPUT 0/20 mA or 4/20 mA

Select the right scale 4...20 mA or 0...20 mA for output signal, then confirm.

SAVE ?

To record the settings, press the icon SAVE

9.6

FORCED RELAY

This menu allows to test manually the relays S1, S2, S3 and S4 through simulations.

Status of last one (S4) is OFF (non actuated) by default. Test begins with threshold S1 with status relay OFF.
To test it, press the arrows SELECTION to switch relay to ON status (activated).
Confirm to go to next threshold; Same routine for testing S2, S3 and S4.

MES

23-11-2018

M-107.01-EN-AE

107-01

/12

pH/ mV monitors

BAMOPHAR 107

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Summary of Contents for BAMOPHAR 107

Page 1: ...MANUAL MES 23 11 2018 M 107 01 EN AE 107 01 1 pH mV monitors BAMOPHAR 107 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www bamo...

Page 2: ...RELAY OUTPUTS S1 S2 AND S3 AS THRESHOLDS 9 2 1 11 COMMENTS ABOUT REGULATION WITH BAMOPHAR 9 2 2 11 ADJUSTMENT OF THRESHOLDS S1 S2 AND S3 IN MODE ON OFF REGULATION 9 2 3 11 ASSIGNMENT OF S3 THRESHOLD...

Page 3: ...sensor or cleaning mode Contact Initial resistance 100 m max voltage drop 6 V DC 1 A Switching power 831 VA AC 3 A 277 V AC 90 W 3 A 30 V DC Switching capacity min 100 mA 5 V DC depending of frequency...

Page 4: ...top cover Measurement input Screw connector for pH cable type 9060 Note Signal input on screw connector for 9060 cable type Only for wall mount models codes 107520 107521 107523 and 107524 MES 23 11...

Page 5: ...O contact for measurement temperature 21 S1 22 Relay S3 N O contact for measurement or temperature 23 S3 May be assigned to the external signal becomes NC contact see 9 2 3 24 25 Main power supply 26...

Page 6: ...ayed as pH 7 4 3 1 9060 coaxial cable preparation 1 4 2 5 6 3 Remove the outer insulation by 20 mm Fold back the shield copper mesh Remove the 2nd sheath black layer over 20 mm Strip the core of the c...

Page 7: ...ontact between the copper shield and the 2 brackets will be perfect 6 Press on the cable to insert it properly in the brackets 7 Screw on tightly to block the tip in its connector 8 Replace the lid of...

Page 8: ...tatus of relays S1 S2 S3 S4 Designations Identifications of your instruments Measurements display Parameter settings of both BAMOPHAR Display settings Date and time Measurements details Opened or clos...

Page 9: ...Confirm setting and go to the next parameter SAVE To save the parameters Closed means CONSULTATION review mode e g Green icon for extension unit 7 DISPLAY SETTINGS Note To access to the display setti...

Page 10: ...buffer pH 4 00 for a process with acids Press MENU and choose ADJUST ELECTRODE STANDARD 7 00 pH Plunge the tip of the electrode in the pH 7 00 buffer solution Ref 9011 then confirm Do not forget to re...

Page 11: ...e the relay S1 select OFF mode then confirm and save the setting ALARM 1 MEASURE TEMP In ON mode threshold S1 can be set in MEASUREMENT or TEMPERATURE mode MEASURE mode threshold dedicated to pH or mV...

Page 12: ...UT 4 20 mA or 0 20 mA Select the right scale 4 20 mA or 0 20 mA for output signal then confirm SAVE To record the settings press the icon SAVE 9 5 SETTINGS OF ANALOGUE OUTPUT FOR TEMPERATURE Go to men...

Page 13: ...tings press the icon SAVE Cycle time High limit Set point Low limit Time S1 actuated No action Dead zone Measurement S2 EXAMPLE For settings such as following Set point pH 7 00 Cycle time 10 s Upper b...

Page 14: ...o go to ADJUST ALARM To access to the CLEANING PROBE menu switch alarms mode to WITHOUT ALARM in menu ADJUST ALARM If the ALARM mode is not activated the CLEANING PROBE function is available CLEANING...

Page 15: ...Ti s Td s PID 1 2 x T t 2 x t 0 5 x t PI 0 9 x T t 3 3 x t 00 P T t 9999 00 PID analogue output 8 4 In MANUAL mode REGUL MANU pH measurement Reached set point Time Time Cde 10 m 10 To To 9 10 TYPE OF...

Reviews: