background image

9.2

SETTING THE RELAY OUTPUTS S1, S2 AND S3 AS THRESHOLDS

9.2.1 COMMENTS ABOUT REGULATION WITH BAMOPHAR

Before to set up or to choose the regulation mode, it is necessary to define the operating mode.

BAMOPHAR 107 allows you to choose between 3 operating modes:

1) ON/ OFF regulation mode with thresholds values S1, S2 and/ or S3
2) Relay regulation mode with proportional bandwidth and dead zones (relays S1 and S2)
3) P.I.D. regulation mode using the 4-20 mA output signal of BAMOPHAR

Specificity of threshold S3:

May be assigned to the external signal (E1): to block the regulation (terminals 5 and 6, see wiring diagram).
In this case, relay S3 becomes a normally closed contact (N.C) and E1 is not active; S3 opens when blocking regulation operates (E1 active).

This means:
- Relay S3 is normally closed (N.C.) if no voltage applies between terminal 5 and 6 (E1 not active).
- Relay S3 will open when voltage applies between terminals 5 and 6 (E1 active)

Consequences of blocking the regulation:

• When working in ON/ OFF regulation mode, there is no additional action on relays S1 and/ or S2.

• However, it is possible for the user to connect the relays S1 and S3 in series to secure the injection (if relay S1 manages the injection of

reagent). When working in relay regulation mode, relays s1 and s2 are opened (non actuated); regulation is inhibited.

• When working in PID regulation mode, the 4-20 mA signal is frozen.

9.2.2 ADJUSTMENT OF THRESHOLDS S1, S2 AND S3 IN MODE ON/OFF REGULATION

Below is a detailed description of the steps to set the threshold S1:

Enter to the menu ADJUST ALARM 1

ALARM 1 ON / OFF

To set up output S1, select ON, then confirm.

To disable the relay S1 select OFF mode, then confirm and save the setting.

ALARM 1 MEASURE / TEMP.

In ON mode, threshold S1 can be set in MEASUREMENT or TEMPERATURE mode:
MEASURE mode = threshold dedicated to pH or mV (ORP) measurements

TEMP. mode = threshold dedicated to temperature measurement
Select the appropriate mode, then confirm.

ALARM 1 HIGH / LOW

HIGH = Trigger when measurement rises over threshold S1
LOW = Trigger when measurement decreases below threshold S1

Select the appropriate actuating mode, then confirm.

ON 08.00 pH / °C

Enter the value at which relay S1 will actuate (closing), then confirm.

OFF 07.50 pH / °C

Enter the value at which relay S1 will be opened, then confirm.

DELAY UP ON / OFF

With or without delay to actuate the relay S1.

TIME 0000 Sec

Enter the duration of delay before actuating S1, then confirm.

DELAY DOWN ON / OFF

With or without delay to open relay S1

TIME 0000 Sec

Enter the duration of delay before opening S1, then confirm.

SAVE ?

To record the settings, press the icon SAVE.

All settings for S2 and S3 are available respectively in ADJUST ALARM 2 and ADJUST ALARM 3 menus, identical to S1.

9.2.3 ASSIGNMENT OF S3 THRESHOLD TO EXTERNAL SIGNAL (E1)

Go to the menu ADJUST ALARM 3.

ADJUST ALARM 3 ON / OFF

Select the ON mode to set up the output S3, then confirm.

EXTERN. YES / NO

If NO : the sequence of settings are same as for S1 or S2 in ON/ OFF regulation mode
If YES = external signal (E1) is assigned to relay S3

Relay S3 becomes a normally closed contact (E1 not active) and will open when blocking of
regulation operates (E1 active).
Confirm the selection.

SAVE ?

To record the settings, press the icon SAVE.

MES

23-11-2018

M-107.01-EN-AE

107-01

/11

pH/ mV monitors

BAMOPHAR 107

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Summary of Contents for BAMOPHAR 107

Page 1: ...MANUAL MES 23 11 2018 M 107 01 EN AE 107 01 1 pH mV monitors BAMOPHAR 107 22 Rue de la Voie des Bans Z I de la gare 95100 ARGENTEUIL Tel Fax 33 0 1 30 25 83 20 33 0 1 34 10 16 05 Web E mail www bamo...

Page 2: ...RELAY OUTPUTS S1 S2 AND S3 AS THRESHOLDS 9 2 1 11 COMMENTS ABOUT REGULATION WITH BAMOPHAR 9 2 2 11 ADJUSTMENT OF THRESHOLDS S1 S2 AND S3 IN MODE ON OFF REGULATION 9 2 3 11 ASSIGNMENT OF S3 THRESHOLD...

Page 3: ...sensor or cleaning mode Contact Initial resistance 100 m max voltage drop 6 V DC 1 A Switching power 831 VA AC 3 A 277 V AC 90 W 3 A 30 V DC Switching capacity min 100 mA 5 V DC depending of frequency...

Page 4: ...top cover Measurement input Screw connector for pH cable type 9060 Note Signal input on screw connector for 9060 cable type Only for wall mount models codes 107520 107521 107523 and 107524 MES 23 11...

Page 5: ...O contact for measurement temperature 21 S1 22 Relay S3 N O contact for measurement or temperature 23 S3 May be assigned to the external signal becomes NC contact see 9 2 3 24 25 Main power supply 26...

Page 6: ...ayed as pH 7 4 3 1 9060 coaxial cable preparation 1 4 2 5 6 3 Remove the outer insulation by 20 mm Fold back the shield copper mesh Remove the 2nd sheath black layer over 20 mm Strip the core of the c...

Page 7: ...ontact between the copper shield and the 2 brackets will be perfect 6 Press on the cable to insert it properly in the brackets 7 Screw on tightly to block the tip in its connector 8 Replace the lid of...

Page 8: ...tatus of relays S1 S2 S3 S4 Designations Identifications of your instruments Measurements display Parameter settings of both BAMOPHAR Display settings Date and time Measurements details Opened or clos...

Page 9: ...Confirm setting and go to the next parameter SAVE To save the parameters Closed means CONSULTATION review mode e g Green icon for extension unit 7 DISPLAY SETTINGS Note To access to the display setti...

Page 10: ...buffer pH 4 00 for a process with acids Press MENU and choose ADJUST ELECTRODE STANDARD 7 00 pH Plunge the tip of the electrode in the pH 7 00 buffer solution Ref 9011 then confirm Do not forget to re...

Page 11: ...e the relay S1 select OFF mode then confirm and save the setting ALARM 1 MEASURE TEMP In ON mode threshold S1 can be set in MEASUREMENT or TEMPERATURE mode MEASURE mode threshold dedicated to pH or mV...

Page 12: ...UT 4 20 mA or 0 20 mA Select the right scale 4 20 mA or 0 20 mA for output signal then confirm SAVE To record the settings press the icon SAVE 9 5 SETTINGS OF ANALOGUE OUTPUT FOR TEMPERATURE Go to men...

Page 13: ...tings press the icon SAVE Cycle time High limit Set point Low limit Time S1 actuated No action Dead zone Measurement S2 EXAMPLE For settings such as following Set point pH 7 00 Cycle time 10 s Upper b...

Page 14: ...o go to ADJUST ALARM To access to the CLEANING PROBE menu switch alarms mode to WITHOUT ALARM in menu ADJUST ALARM If the ALARM mode is not activated the CLEANING PROBE function is available CLEANING...

Page 15: ...Ti s Td s PID 1 2 x T t 2 x t 0 5 x t PI 0 9 x T t 3 3 x t 00 P T t 9999 00 PID analogue output 8 4 In MANUAL mode REGUL MANU pH measurement Reached set point Time Time Cde 10 m 10 To To 9 10 TYPE OF...

Reviews: