Table 10
Input Signal, Power, and Display Specifications
Power Supply
Loop powered from 4-20 mA control signal
Compliance Voltage Rating
9.0 V at 20 mA, 11.0 V at 4.0 mA
Minimum Current Signal to Start Up
3.2 mA
Impedance Range
Low: 450 Ohms; High: 2750 Ohms
Minimum Input Span for Split Range Operation
5 mA
Upper Range Value for Split Range Operation
8 mA to 20 mA
Lower Range Value for Split Range Operation
4 mA to 14 mA
Wire Size
14/28 AWG
Strip Length
0.22 in / 6 mm
Digital Communication
HART
®
Communication protocol revision 5, 6 or 7
Local Display
LCD, Explosion proof with two lines of nine alpha
numeric characters.
Display becomes unreadable between 0 °C and -10 °C.
Display is shutdown at -15 °C.
Push Buttons
External, Three Explosion Proof / Flameproof push but-
tons
Masoneilan SVI II AP Digital Positioner Quick Start Guide
=
| 85
© 2020 Baker Hughes Company. All rights reserved
.
Table 11
Construction Material Specifications
Housing and Cover
Aluminum ASTM B85 SG100A standard
Stainless Steel optional
Weight
Standard Flow Model:
Aluminum - 7.4 lbs./ 3.3 kg
Stainless Steel - 16 lbs/ 7.3 kg
High Flow Model:
With Display: 9.4 lbs./ 4.2 kg
Without Display: 8.9 lbs./ 4.0 kg
Relay and Manifold
Standard Flow Model:
Single Acting - PPS, 300 Series Stainless Steel, nitrile diaphragms
Double Acting - 300 Series Stainless Steel, Ryton;
Aluminum 6061 T6, Ryton
High Flow Model:
300 Series Stainless Steel, Ryton;
Aluminum 6061 T6, Ryton
I/P Motor
430 stainless steel, PPS, 300 series stainless steel
Mounting Bracket
300 series stainless steel
Magnet Holder
Corrosion Protected Anodized Aluminum 6061 T6