Copyright 2013 Baker Hughes Company.
English–Dry
TC / Liquid
TC Temperature Calibrators | 43
10.3 Liquid
TC Specification
Specification
Liquid
TC
165
Liquid
TC
255
Calibrator:
Temperature range:
- Micro bath (water)
2
°C to 95
°C
(35.6
°F to 203
°F)
T
R
to 95
°C
(T
R
to 203
°F)
- Micro bath (silicone oil)
-
35
°C to 165
°C
(
-
31
°F to 329
°F)
T
R
to 255
°C
(T
R
to 491
°F)
Accuracy
±
0.1
°C
(±
0.18
°F)
±
0.2
°C
(±
0.36
°F)
Stability
±
0.05
°C
(±
0.09
°F)
Display range.
-
50
°C to 165
°C
(
-
58
°F to 329
°F)
0
°C to 255
°C
(32
°F to 491
°F)
Electrical Characteristics:
Power supply:
- Voltage range.
100
Vac to 240
Vac
100
Vac to 230
Vac
- Frequency
50 / 60
Hz
Power consumption.
375
W
1000
W
Fuses
2 off 5 × 20
mm
T6.3H250V
2 off 5 × 20
mm
T10H250V
Serial interface:
- Interface type.
RS
485 optoisolated
- Communication protocol.
MODBUS RTU (JBUS)
- Serial transfer speed.
9600 Baud
Dimensions:
Dry block:
- Bore diameter
60
mm (2.36″)
- Depth
170
mm (6.69″)
- Measurement zone.
110
mm to 150
mm (4.33″ to 5.91″)
- Sensor cage working depth.
150
mm (5.91″)
Housing dimensions:
- Width
210
mm (8.27″)
150
mm (5.91″)
- Handle
380+50
mm (15.0+2.0″)
330+70
mm (13.0+2.8″)
- Depth
300
mm (11.8″)
270
mm (10.6″)
- Weight
~
12.5
kg (~
27.6
lbs)
~
7.5
kg (~
16.5
lbs)