Copyright 2013 Baker Hughes Company.
12 | Dry
TC / Liquid
TC Temperature Calibrators–English
The calibration liquid is poured directly into the micro bath tank / tub insert.
The micro bath includes the following accessories:
•
Transport Cover
•
Work Cover
•
Sensor Cage
•
Magnetic Stirrer
•
Drain Syringe
•
Magnetic Lifter
Note:
A removable tub insert is available as an optional accessory.
The accessories are described in more detail in the following sections.
4.3.5.1 Transport Cover
The transport cover seals the micro bath, preventing spillage of the calibration liquid during
transportation.
Figure
7: Transport Cover
4.3.5.2 Work Cover
Figure
8: Work Cover
The work cover is used to:
a. Reduce evaporation of the calibration liquid.
b. Reduce the cooling of the surface of the calibration liquid.
c.
Provide stable positioning of the DUT in the micro bath.
WARNING
Remove the transport cover before using the micro bath.
The transport cover has a pressure relief safety valve. The valve opens when the
micro bath pressures exceeds ~1.5
bar. This can result in hot vapor being
released.