Copyright 2013 Baker Hughes Company.
English–Dry
TC / Liquid
TC Temperature Calibrators | 31
See Table
5 for the heating and cooling capacity of the Dry
TC and Liquid
TC calibrators.
5.4.4.1 Heating and Cooling Gradient Parameter Level
The heating and cooling gradients and duration time can be set in parameter level
(regulation).
1. Press the
key for approximately 5
seconds. The main menu opens.
The last selected function appears on the upper display.
LED SET flashes on the lower display.
2. Press the
or
key until
(operator) is displayed.
3. Press the
key again. The group level opens.
Figure
37: Group Level SP (Set Point)
appears on the upper display.
Table
5: Calibrator Maximum Heating and Cooling Capacity
Model
Heating Capacity
Cooling Capacity
Dry
TC
165
7
°C/min
(12.6
°F/min)
5
°C/min
(9
°F/min)
Dry
TC
650
35
°C/min
(63
°F/min)
[650
°C to 300
°C]
(1202
°F to 572
°F)
10
°C/min
(18
°F/min)
[300
°C to 100
°C]
(572
°F to 212
°F)
5
°C/min
(9
°F/min)
Liquid
TC
165
3
°C/min
a
(5.4
°F/min)
a. Silicone oil 10
cSt.
6
°C/min
a
(10.8
°F/min)
5
°C/min
b
(9
°F/min)
b. Distilled water.
4
°C/min
b
(7.2
°F/min)
Liquid
TC
255
22
°C/min
c
(39.6
°F/min)
c. Silicone oil 50
cSt.
[200
°C to 50
°C]
(392
°F to 122
°F)
4
°C/min
c
(7.2
°F/min)
(50
°C to 30
°C)
(122
°F to 86
°F)
0.5
°C/min
c
(0.9
°F/min)
12
°C/min
b
(21.6
°F/min)
(90
°C to 50
°C)
(194
°F to 122
°F)
2
°C/min
b
(3.6
°F/min)
(50
°C to 30
°C)
(122
°F to 86
°F)
0.5
°C/min
b
(0.9
°F/min)