32 | Dry
TC / Liquid
TC Temperature Calibrators–English
(set point) appears on the lower display and LED SET flashes.
4. Press the
key to select group level
.
Figure
38: Group Level ‘rEG (Regulation)
appears on the upper display.
appears on the lower display and LED SET flashes.
5. Press the
key again. The parameter level opens.
Figure
39: Parameter Level SLor (Heating Gradient)
appears on the upper display.
flashes on the lower display.
5.4.4.2 Setting the Heating Gradient
The heating gradient
is active if the reference temperature is lower than the set point
temperature.
The heating gradient is adjustable from 99.99
°C/min to 0.00
°C/min. See Table
5 for the
maximum heating capacity of the calibrator.
Note:
If
=
(in no function), heating gradient control has been disabled.
1. Enter the gradient control parameter level. See Section
Figure
40: Parameter Level SLor (Heating Gradient)
appears on the upper display.
flashes on the lower display.