Copyright 2013 Baker Hughes Company.
English–Dry
TC / Liquid
TC Temperature Calibrators | 17
2. Place the magnetic stirrer into the micro bath.
3. Insert the sensor cage.
4. Insert the DUT into the sensor cage, before filling, to account for its volume.
5. Fill the micro bath with calibration liquid.
Note:
Observe the maximum fill level and leave sufficient reserve space for an additional
rise in the level.
6. If necessary, remove the DUT.
7. Screw the work cover into place.
8. Insert the DUT through the work cover into the micro bath.
Tub insert filling procedure:
1. Unscrew the transport cover of the tub insert.
2. Insert the tub insert into the micro bath tank using the sleeve exchange tool.
3. Place the magnetic stirrer into the tub insert.
4. Insert the sensor cage.
5. Insert the DUT into the sensor cage, before filling, to account for its volume.
6. Fill the tub insert with calibration liquid.
Note:
Observe the maximum fill level and leave sufficient reserve space for an additional
rise in the level.
7. If necessary, remove the DUT.
8. Screw the work cover into place.
9. Insert the DUT through the work cover into the tub insert.
4.4 Switching On, Cool Down and Switching Off
For reasons of safety, when switching on the calibrator, the fan runs at the fastest speed. As soon
as the internal reference sensor has measured a safe temperature, the fan speed is adjusted
accordingly.