Copyright 2013 Baker Hughes Company.
English–Dry
TC / Liquid
TC Temperature Calibrators | 23
a.
indicates the heating capacity.
b.
indicates the cooling capacity.
5.2.4.2 Operating Duration
1. Briefly press the
key.
2. The current number of operating hours of the calibrator is displayed for approximately 5
s.
5.2.5 Configuration Menu
The configuration menu contains important system parameters. These parameters shall only be
modified by an authorized Druck service centre.
5.3 Temporary Set Point Mode
In this operating mode, it is possible to temporarily modify a saved set point temperature.
Figure
20: Temporary Set Point Mode
1. Press the
key briefly.
a. The current active set point, e.g.
(set point 2), is displayed on the upper
display.
b. The respective set point temperature is displayed by the lower display.
2. Press the
key to increase the set point temperature.
3. Press the
key to decrease the set point temperature.
4. Press the
key again to confirm the new set point.
Note:
Press the
and
key to raise and lower the value by 0.1 respectively. If the keys are
held pressed for at least one second, the value increases or decreases quickly and after two
seconds even more quickly; this means the desired value can be reached rapidly.
Note:
If no key is pressed for approximately 15
seconds, the controller automatically returns to
the calibration mode.
5.4 Main Menu
All settings can be found in the main menu.
1. Press the
key for approximately 5
seconds. The main menu opens.
2. Use the
and
keys to select the desired main menu, see Figure
INFORMATION
The
(configuration) menu is protected by a
password.
Unauthorized access can damage the calibrator.
Exit the menu by pressing the
key twice.