Copyright 2013 Baker Hughes Company.
English–Dry
TC / Liquid
TC Temperature Calibrators | 27
3. Press the
key to confirm.
Figure
27: Manual Control (OPLO) Display
The current reference temperature appears on the upper display.
The letter and the current percentage of heating capacity appear on the lower display.
4. Press the
key to increase the output capacity.
5. Press the
key to decrease the output capacity.
Note:
Press the
and
key to raise and lower the value by 0.1 respectively. If the keys are
held pressed for at least one second, the value increases or decreases quickly and after two
seconds even more quickly; this means the desired value can be reached rapidly.
5.4.2.2 Switching Off Manual Control
Manual control is switched on if the following display appears:
Figure
28: Manual Control (OPLO) Display
The current reference temperature appears on the upper display.
The letter and the current percentage of heating capacity appear on the lower display.
To switch off manual control:
1. Press the
key for approximately 5
seconds. The main menu opens.
Figure
29: Manual Control (OPLO) Menu
appears on the upper display.
LED SET flashes on the lower display.
2. Press the
or
key until the desired operating mode is displayed.
3. Press the
key to confirm.