H25-IR PRO Instruction Manual
Operation
P/N: 3015-5678 Rev 6
51
NOTE:
If you set the alarm setpoint to a non-zero value, the alarm “beep” sounds only when the gas sample
exceeds the alarm setpoint. If you set the alarm setpoint to zero, the alarm will sound for
any
detected gas, and
will increase in frequency based on the concentration of the gas.
5.5.4. Display Style
The Display Style setting defines the format of the main Search screen. Options are Graphic and Numeric. You choose your
preference from the Select Display Style screen.
•
Graphic
o
Bar graph and basic setup information
o
The Alarm Setpoint setting is used to establish the bar graph’s mid-point and scaling
o
Default setting
•
Numeric
o
Large digital numeric display of the real-time gas concentration and basic setup information
NOTE:
Display Style only affects the
Search
mode display. Measure mode
always
shows a numeric display.
NOTE:
The dark area shown in the bar graph is a real time display of the detected gas. The bar fills to the right
as the gas level increases. When the alarm setpoint is reached or exceeded the speaker sounds.
Figure 5-9. Samples of Graphic and Numeric Display Styles
GAS TYPE = R134A
0
20.0
ALARM >= 10.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
Summary of Contents for H25-IR PRO
Page 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Page 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Page 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Page 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Page 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...