Operation
H25-IR PRO Instruction Manual
68
P/N: 3015-5678 Rev 6
5.6.6. Loop Factor
NOTE:
The Loop Factor setting configures the 4-20mA current output for the sample range. By default, the
standard 4-20mA output is mapped linearly to the input reading range (for example, with “ppm” set as the unit
of measure, and a range of 0 ppm to 1000 ppm, each of the 1000 ppm steps corresponds to the default
0.016 mA). This can be adjusted to your particular application using the Loop Factor setting and the following
equation.
𝐋𝐨𝐨𝐩 𝐎𝐮𝐭𝐩𝐮𝐭 𝐢𝐧 𝐦𝐀 = (𝐋𝐨𝐨𝐩 𝐅𝐚𝐜𝐭𝐨𝐫 × 𝐏𝐏𝐌) + 𝟒𝐦𝐀
Table 5-25. Default Loop Factor Mapping Example (0.016 mA = 1 PPM)
Output
PPM (Default)
Output
PPM (Default)
4 mA DC
0 PPM
13 mA DC
563 PPM
5 mA DC
63 PPM
14 mA DC
625 PPM
6 mA DC
125 PPM
15 mA DC
688 PPM
7 mA DC
188 PPM
16 mA DC
750 PPM
8 mA DC
250 PPM
17 mA DC
813 PPM
9 mA DC
313 PPM
18 mA DC
875 PPM
10 mA DC
375 PPM
19 mA DC
937 PPM
11 mA DC
438 PPM
20 mA DC
1000 PPM
12 mA DC
500 PPM
Table 5-26. Changing the Loop Factor
Step
Button
Description
Sample Screen(s)
1
See Table 5-14
on page 53.
Follow the steps in Table 5-14
on page 53 to highlight the
Loop Factor option.
<<
ADVANCED SETUP
>>
DATE/TIME SET
LOG LIMIT
LOOP FACTOR
SERIAL DATA
PROBE BUTTON
qp
SELECT
± EDIT
Summary of Contents for H25-IR PRO
Page 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Page 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Page 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Page 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Page 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...