H25-IR PRO Instruction Manual
Operation
P/N: 3015-5678 Rev 6
41
5.3.7. Zero (ENT) Button
The ZERO button manually resets the current gas level to zero. It is also used as the ENT button when
changing configuration values within the menus.
NOTE:
Automatic zeroing occurs at startup during initialization. It is therefore
recommended to perform startup of the H25-IR PRO in a refrigerant-free clean-air
environment.
NOTE:
In Search Mode, the H25-IR PRO performs an auto zero process to remove the
effects of background refrigerant in the air. Measure Mode does
not
automatically zero.
Figure 5-5 and Figure 5-6 show a sample Search screen and Measure screen both before and shortly
after the ZERO button is pressed.
Figure 5-5. Sample Search and Measure Screens Before Zeroing
Figure 5-6. Sample Search and Measure Screens Shortly After Zeroing
5.3.8. Information (
i
) Button
The Information (
i
) button provides a shortcut to the DIAGNOSTICS screen. For more information
on the Diagnostics screen, refer to the Diagnostics Screen section on page 74. Faults and other
diagnostics are displayed in the header portion of the main screen. The
i
symbol is shown along
with any fault condition to prompt the user to use the
i
keypad button.
NOTE:
The Information button is also used to setup a leak reference. For more information, refer to the sections
on Adjustment to a Known External Leak Source (on page 81) and Using a Reference Leak Source (on page 6).
GAS TYPE = R134A
0
20.0
ALARM >= 10.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
GAS TYPE = R134A
0
20.0
ALARM >= 10.0 G/YR
SEARCH {---------}
GAS TYPE = R134A
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
Summary of Contents for H25-IR PRO
Page 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Page 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Page 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Page 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Page 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...