Operation
H25-IR PRO Instruction Manual
74
P/N: 3015-5678 Rev 6
5.7. Diagnostics Screen
Table 5-31. Accessing the Diagnostics Screen (Method #1)
Step
Button
Description
Sample Screen(s)
1
Access the main Search or
Measure screen by pressing
the ESC button.
Check the header for a fault
notification.
Note that multiple faults may
exist, but only the first is
displayed.
2
Press the Info (
i
) button
when the
i
icon appears in
the header of the main Search
or Measure screen. The
Current Faults option of the
Diagnostics menu is
highlighted.
3
Press the ENT button to select
the Current Faults option and
display the first fault.
4
Press the Right Arrow button
to view the next fault (if one
exists).
i
0810
:
PRESSURE
OUT
OF
RANGE
ALARM
>=
10.0
RANGE = 0-20 G/YR
MEASURE {---------}
<<
DIAGNOSTICS
CURRENT FAULTS
0810
SENSOR DATA
qp
SELECT
± EDIT
FAULT 0800 >>
PRESSURE TRANSDUCER OUTPUT
IS OUT OF RANGE. CHECK
TUBING.
<<
FAULT 0010
ERROR 0010: 4-20 MA
CURRENT LOOP IS OPEN,
OR CIRCUIT HAS HIGH
RESISTANCE.
CHECK JUMPER OR WIRES
¯
Summary of Contents for H25-IR PRO
Page 6: ...H25 IR PRO Instruction Manual vi P N 3015 5678 Rev 6 ...
Page 8: ...H25 IR PRO Instruction Manual viii P N 3015 5678 Rev 6 ...
Page 39: ...H25 IR PRO Instruction Manual Menu System Overview P N 3015 5678 Rev 6 29 ...
Page 109: ...H25 IR PRO Instruction Manual Maintenance P N 3015 5678 Rev 6 99 ...
Page 120: ...H25 IR PRO Instruction Manual Declaration of Conformity P N 3015 5678 Rev 6 110 ...