Operation
H25-IR PRO Instruction Manual
86
P/N: 3015-5678 Rev 6
Step
Button
Description
Sample Screen(s)
6
Press the SETUP (Right Arrow)
button again to access the
Diagnostics screen.
(Notice the right arrows in the
header, indicating an additional
menu.)
7
Press the SETUP (Right Arrow)
button again to access the Factory
Setup screen.
8
Press the RANGE (Down Arrow)
button six (6) times to highlight
the Factory Gas Type option.
9
Press the ENT button to select the
Gas Type option and to display the
Factory Gas Type screen.
<<
DIAGNOSTICS
>>
CURRENT FAULTS
0800
SENSOR DATA
qp
SELECT
± EDIT
<<
FACTORY SETUP
EMITTER ADJ
DETECTOR ADJ
MASS FLOW
LCD CONTRAST
PPM GAS CAL
¯
qp
SELECT
± EDIT
<<
FACTORY SETUP
DETECTOR ADJ
MASS FLOW
LCD CONTRAST
PPM GAS CAL
GAS TYPE
¯
qp
SELECT
± EDIT
FACTORY GAS TYPE
p
---
DISABLE
q
R21
± SELECT
ESC=QUIT