Instruction 24-9395
50
Fyrite Pro
5.10 CO Sensor Replacement
(Model 100/105/120/125)
Be sure to perform all of the checks lsted n Secton 4.13.13 to ensure that
the CO sensor needs to be replaced.
Material Required:
• CO Sensor, P/N 24-7265
• CO Sensor Gasket, P/N 24-1112
• #1 Phllps Screwdrver
Procedure:
1. Dsassemble the analyzer as follows:
a. Remove the battery cover and the batteres, uncoverng one of
the cover hold-down screws.
b. Remove and set asde all four cover hold-down screws.
c. Wth the analyzer on ts back, remove the front cover, layng t
face down to the left of the body.
d. Carefully remove the crcut board, slppng off the battery con-
nector on top, and then layng the crcut board face down n the
top cover.
2. Gently pull CO sensor out of ts socket (see Fgure 5-5).
3. Properly dspose of the old CO sensor (see the nstructon sheet that
comes wth the new sensor).
4. It s recommended that the CO sensor gasket be replaced at the
same tme as the sensor. Remove the old gasket and dscard. Re-
move the paper backng from the new gasket and adhere t to the
case n the same poston as the old one.
Important:
Ensure that the new gasket does not restrict the
inlet and exhaust holes in the case.
5. Plug the new CO sensor nto ts socket.
6. Reassemble the analyzer.