Instruction 24-9395
42
Fyrite Pro
5.2 Ambient Temperature Calibration
(Model 110/120/125)
Material Required:
Calbrated Thermometer
Procedure:
1. Enter the Calbraton Mode as descrbed n Secton 5.1. Observe
that “AMBIENT” wll appear at the top of the dsplay; f not, re-
peatedly press
eNTeR
untl “AMBIENT” s dsplayed.
2. Use the
and
buttons to set the dsplayed value to match the
readng of a calbrated thermometer at room temperature.
3. Press
eNTeR
to store the dsplayed value and move to the next
calbraton screen, or hold down
eNTeR
for 2 seconds to store the
dsplayed value and enter the Run Mode, or press the
I/O
button to
ext the Calbraton Mode and turn OFF the analyzer wthout sav-
ng the changes.
5.3 Combustion Air Thermocouple
Calibration
(Model 125)
Material Required:
Thermocouple Smulator
Procedure:
1. Attach the thermocouple smulator to the Combuston Ar Thermo-
couple connector on the analyzer (T-AIR, see Fgure 3-2).
2. Enter the Calbraton Mode as descrbed n Secton 5.1. Then re-
peatedly press the
eNTeR
button untl “TA” appears at the top of
the dsplay.