Instruction 24-9395
12
Fyrite Pro
3.5 Setup Mode
The analyzer s preset at the factory for the parameters shown below, but
can be changed as descrbed n ther assocated sectons.
Function
Parameter
To Change
Fuel
Natural Gas (F1)
Secton 4.13.5
Temperature Unt
°F
Secton 3.5.2
CO Channel
Auto Zero (A2)
Secton 3.5.3
Draft (Pressure)
WC
Secton 3.5.4
Clock
Not Intalzed
Secton 3.5.5 thru 3.5.7
Prnter Protocol
IrDA
Secton 3.5.8
3.5.1 Entering Setup
1. Wth the analyzer turned OFF, press and hold down the
hOlD
button.
2. Press the
I/O
button.
3. Release both buttons.
4. The analyzer s now n ts Setup Mode. Refer to Sectons 3.5.2 thru
3.5.8 for nformaton on how to set up each analyzer parameter.
3.5.2 Selecting Temperature Units
The Temperature Setup Screen s labeled “Unt.”
1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly
press the
eNTeR
button untl “Unt” s dsplayed.
2. Press the
or
button untl the desred temperature unt (°F or
°C) s dsplayed.
3. Press
eNTeR
to move to the next Setup Screen, or
I/O
to ext setup.