5 Test Procedure
46
DHV0078 b2 EN
©
b2 electronic
GmbH
5.4
Interrupting a Test
Once a test has started, it can be interrupted at any time. It is recommended to
select the appropriate method corresponding to the situation.
Situation
Procedure
Example
Routine STOP
(No emergency)
When a test is in progress, “STOP” on the display screen
is highlighted.
To interrupt the test, push in/click the navigation knob
.
•
HVA
software
deactivates HV
•
Test stops
Alternative
When a test is in progress, press the HV switch
to
deactivate high voltage.
•
HVA
hardware
deactivates HV
•
Test stops
Emergency stop
In an emergency situation, press the emergency OFF
to shut down the system.
•
HVA
hardware
deactivates HV
•
Test stops
After test interruption, a message is displayed indicating that the test has been
terminated by the user:
Manual Mode Sine 0.1Hz
STOP
T: 00:58 / 1min
1.0
kV
rms
0.1kV
0.9µA
0.4nF
13GΩ
Manual Mode Finished
Manual Mode Test Seq.
interrupted by the user at
step 1 after 58s
Sine 0.1 Hz
Test Voltage: 1.0 kVrms
OK
Summary of Contents for HVA120
Page 2: ...1 Introduction 2 DHV0078 b2 EN b2 electronic GmbH...
Page 11: ...2 Safety b2 electronic GmbH DHV0078 b2 EN 11...
Page 33: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 33 Connection diagram vacuum bottle test...
Page 47: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 47 5 5 Discharge Status...
Page 63: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 63 HVA30 5...
Page 64: ...11 Declaration of Conformity 64 DHV0078 b2 EN b2 electronic GmbH HVA30 7 HVA40 5 HVA54 3...
Page 65: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 65 HVA54 5...
Page 66: ...11 Declaration of Conformity 66 DHV0078 b2 EN b2 electronic GmbH HVA60...
Page 67: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 67 HVA68 2...
Page 68: ...11 Declaration of Conformity 68 DHV0078 b2 EN b2 electronic GmbH HVA90 HVA94...
Page 69: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 69 HVA120...