7 Disconnection Procedure
54
DHV0078 b2 EN
©
b2 electronic
GmbH
7
Disconnection Procedure
Disconnection diagram
Normal conditions
Steps
D1-D8
describe the
normal disconnection
procedure.
Step
Procedure
(normal disconnection)
D1
Press Emergency OFF
Stop test according 5.4. and press emergency OFF to lock against re-energizing
D2
Verify HV status
•
Wait until LED red
light deactivates (indicates residual voltage < 100 V)
D3
Discharge and earth the DUT complying with local safety regulations
D4
Turn OFF HVA
•
Turn the HVA main switch
off
D5
Lock HVA in disabled state to prevent unauthorized use:
•
Turn key switch
to OFF position and remove key
DANGER
Electric shock hazard
Never assume that equipment is safe to handle without using the required safety
equipment and earthing procedures.
Disconnection procedures must comply with local safety
regulations.
•
Before disconnecting test lead, DUT must be discharged and earthed.
•
Earth connections must be removed last!
Summary of Contents for HVA120
Page 2: ...1 Introduction 2 DHV0078 b2 EN b2 electronic GmbH...
Page 11: ...2 Safety b2 electronic GmbH DHV0078 b2 EN 11...
Page 33: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 33 Connection diagram vacuum bottle test...
Page 47: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 47 5 5 Discharge Status...
Page 63: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 63 HVA30 5...
Page 64: ...11 Declaration of Conformity 64 DHV0078 b2 EN b2 electronic GmbH HVA30 7 HVA40 5 HVA54 3...
Page 65: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 65 HVA54 5...
Page 66: ...11 Declaration of Conformity 66 DHV0078 b2 EN b2 electronic GmbH HVA60...
Page 67: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 67 HVA68 2...
Page 68: ...11 Declaration of Conformity 68 DHV0078 b2 EN b2 electronic GmbH HVA90 HVA94...
Page 69: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 69 HVA120...