3 General Description
14
DHV0078 b2 EN
©
b2 electronic
GmbH
Characteristic
1
HVA54-5
HVA90
HVA94
HVA120
Article number
SH5013
SH5017
SH5018
SH5019
Input supply voltage
190-240 V; 50/60
Hz
; ±10%
Input supply power
6 kVA
3 kVA
max. output
voltage
VLF sine wave
38 kV
rms
54 kV
peak
64 kV
rms
90 kV
peak
66 kV
rms
94 kV
peak
85 kV
rms
120 kV
peak
VLF square wave
54 kV
90 kV
100 kV
DC [+/-]
54 kV
90 kV
100 kV
resolution: 0.1 kV, accuracy: ± 1%
max. output current
120 mA
ms
41 mA
rms
41 mA
rms
56 mA
rms
resolution: 1 µA, accuracy: ± 1%
Resistance range
0.1 MΩ….5 GΩ
Output frequency
0.01 Hz-0.1 Hz in steps of 0.01 Hz (default 0.1 Hz) –
auto frequency selection
Sheath test
max. test voltage: 10 kV, duration: 1 min-15 min
trip current: 0.1 mA-5.0 mA
Sheath fault location mode
2
max. test voltage: 10 kV, duration: 1 min-60 min
pulse/period: 1:3 / 4 s, 1:5 / 4 s, 1:5 / 6 s, 1:9 / 6 s
Frequency optimization
Yes
Output load
capacity
at 0.1 Hz
5 µF
1 µF
0.85 µF
1 µF
at reduced
frequency/voltage
12 µF
10 µF
10 µF
5 µF
Metering
voltage and current (true rms and/or peak),
capacitance, resistance, time, flashover voltage
Output duty
Continuous! No thermal limitation for operating time.
Test modes
manual & automatic
Output modes
AC (VLF) symmetrical and load independent across full range,
DC (plus or negative polarity), burn/fault condition or
fault trip mode, jacket/sheath testing
Safety
50 Hz-12 kV feedback protection/dual discharge device (internal)
Computer interface
RS232 & USB flash memory module
Record storage
Built in memory: up to 50 reports, 40 test sequences
USB memory flash drive: unlimited
PC software [included]
b2 ControlCenter for Windows
Weight
169 kg
127 kg
128 kg
198 kg
Dimensions
4
(mm) L x W x H
863 x 445 x 610
545 x 445 x 610
790 x 445 x 740
Environment
Storage temp.
-25°C to 70°C
Operation temp.
-10°C to 50°C
Humidity
5-85% non-condensing
1
Technical Specifications are subject to change. b2 reserves the right to modify values in accordance with future HVA
development
.
2
in combination with locator set (not in scope of supply) |
3
50 Hz-12 kV feedback protection (optional)
4
dimensions without handles
Summary of Contents for HVA120
Page 2: ...1 Introduction 2 DHV0078 b2 EN b2 electronic GmbH...
Page 11: ...2 Safety b2 electronic GmbH DHV0078 b2 EN 11...
Page 33: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 33 Connection diagram vacuum bottle test...
Page 47: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 47 5 5 Discharge Status...
Page 63: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 63 HVA30 5...
Page 64: ...11 Declaration of Conformity 64 DHV0078 b2 EN b2 electronic GmbH HVA30 7 HVA40 5 HVA54 3...
Page 65: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 65 HVA54 5...
Page 66: ...11 Declaration of Conformity 66 DHV0078 b2 EN b2 electronic GmbH HVA60...
Page 67: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 67 HVA68 2...
Page 68: ...11 Declaration of Conformity 68 DHV0078 b2 EN b2 electronic GmbH HVA90 HVA94...
Page 69: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 69 HVA120...