Characteristic
1
HVA30-7
HVA40-5
HVA54-3
HVA68-2
Article number
SH5005
SH5009
SH5012
SH5016
Input supply voltage
190-240 V, 50/60 Hz
Input supply power
3 kVA
max. output
voltage
VLF sine wave
24 kV
rms
34 kV
peak
32 kV
rms
45 kV
peak
38 kV
rms
54 kV
peak
48 kV
rms
68 kV
peak
VLF square wave
34 kV
45 kV
54 kV
60 kV
DC [+/-]
34 kV
45 kV
54 kV
60 kV
resolution: 0.1 kV, accuracy: ± 1%
max. output current
120 mA
rms
52 mA
rms
resolution: 1 µA, accuracy: ± 1%
Resistance range
0.1 MΩ….5 GΩ
Output frequency
0.01 Hz-0.1 Hz in steps of 0.01 Hz (default 0.1 Hz) –
auto frequency selection
Sheath test
max. test voltage: 10 kV, duration: 1 min-15 min
trip current: 0.1 mA-5.0 mA
Sheath fault location mode
2
max. test voltage: 10 kV, duration: 1 min-60 min
pulse/period: 1:3 / 4 s, 1:5 / 4 s, 1:5 / 6 s, 1:9 / 6 s
Frequency optimization
Yes
Output Load
Capacity
at 0.1 Hz
7 µF
5 µF
3 µF
2 µF
at reduced
frequency/voltage
15 µF
15 µF
10 µF
10 µF
Metering
voltage and current (true rms and/or peak),
capacitance, resistance, time, flashover voltage
Output duty
Continuous! No thermal limitation for operating time.
Test modes
manual & automatic
Output modes
AC (VLF) symmetrical and load independent across full range,
DC (plus or negative polarity), burn/fault condition or
fault trip mode, jacket/sheath testing
Safety
50 Hz-12 kV feedback protection/dual discharge device (internal)
Computer interface
RS232 & USB flash memory module
Record storage
Built-in memory: up to 50 reports, 40 test sequences
USB memory flash drive: unlimited
PC software [included]
b2 ControlCenter for Windows
Weight
57 kg
Dimensions
4
(mm) L x W x H
450 x 340 x 520
Environment
Storage temp.
-25°C to 70°C
Operation temp.
-10°C to 50°C
Humidity
5-85% non-condensing
Summary of Contents for HVA120
Page 2: ...1 Introduction 2 DHV0078 b2 EN b2 electronic GmbH...
Page 11: ...2 Safety b2 electronic GmbH DHV0078 b2 EN 11...
Page 33: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 33 Connection diagram vacuum bottle test...
Page 47: ...5 Test Procedure b2 electronic GmbH DHV0078 b2 EN 47 5 5 Discharge Status...
Page 63: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 63 HVA30 5...
Page 64: ...11 Declaration of Conformity 64 DHV0078 b2 EN b2 electronic GmbH HVA30 7 HVA40 5 HVA54 3...
Page 65: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 65 HVA54 5...
Page 66: ...11 Declaration of Conformity 66 DHV0078 b2 EN b2 electronic GmbH HVA60...
Page 67: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 67 HVA68 2...
Page 68: ...11 Declaration of Conformity 68 DHV0078 b2 EN b2 electronic GmbH HVA90 HVA94...
Page 69: ...11 Declaration of Conformity b2 electronic GmbH DHV0078 b2 EN 69 HVA120...