Summary of Contents for WorldWEIGH C100 Series

Page 1: ...C100 Counting Scale Technical Manual V2 40...

Page 2: ......

Page 3: ...e to determine the Unit Weight 11 Enter a known Weight 12 Automatic update of Unit Weight 12 Check weighing of count pre set 12 Setting the checking limits 12 PLU Product Look Up 13 Storing PLU 13 Ent...

Page 4: ...b 00005 kg 0001 kg 0002 kg 0005 kg Units of Measure lb kg Interface Bi directional RS 232 Interface Stabilization Time 2 Seconds Operating Temperature 32 F 104 F Power supply AC Adaptor 9 V 800 mA Bat...

Page 5: ...air conditioning vents Avoid unsuitable tables The table or floor must be rigid and not vibrate Do not place near vibrating machinery Avoid unstable power sources Do not use near large users of electr...

Page 6: ...is not damaged and that all accessories are included If equipped remove the shipping safety screw Indicated by the red label on the bottom of the C100 Accessories 1 C100 2 Power Adaptor 3 Stainless st...

Page 7: ...ld damage the load cell inside Check the water mark If bubble is not centered adjust the leveling feet until centered Check the level when you change the location Attach the power supply cable to the...

Page 8: ...Press key to change between remote base and local base Remote Scale Connection The cable from the remote base goes to a 9 pin male DB9 connecter on the left side of the C100 with following connections...

Page 9: ...the total weight and displays the result as a net weight Manually enter the weight of sample also changes the unit from lb to kg if they are enabled To enter the numbers of items used as the sample un...

Page 10: ...hing Mode Unit Weight Display Smpl of sample is very low U Wt Unit weight is below the minimum weight M Data entered into the memory Local Remote Active Scale in use Count Display CkPcs Active in Coun...

Page 11: ...ghing functions are same for both the scales local or remote The number of weighing divisions may be less on the remote scale dependant on the total capacity of the load cell s used 3 Zero To zero the...

Page 12: ...a tare value of 103g is entered into the 60Kg scale with 5g readability then the display will be shown 105g 5 Accumulation The C100 can totalize weight values or count quantities Manual Accumulation...

Page 13: ...for display stability and a control beep Unload the item from the platform the weighing value is added into the memory The scale must return to come zero or negative number before adding another sampl...

Page 14: ...ount will increase If the scale is not stable the calculation will not be completed If the weight is below zero Count display will show negative count Enter a known Unit Weight If an a known unit weig...

Page 15: ...ect counting Press key the active high count limit will be shown Use numeric key to enter desired and if necessary press to clear Press key the active low count limit will be shown Use numeric key to...

Page 16: ...ht to be stored can be either taken from a weighing in process or by enter manually Press display will be shown Press numeric key exp 27 Press Currently stored text will appear The first digit is flas...

Page 17: ...8 T U V 9 W X Y Z 0 _ _ space Note that this method is only used where alpha numeric data is permitted This is used for the Description field and the User ID number Scale ID number in the parameters...

Page 18: ...splay will be show Select the Menu Block Press key it can choose menu block one by one Enter the Selected Menu Press key to confirm what the displays shows Select the Sub Menu Press key it can choose...

Page 19: ...sable 3 Scale will turn off after 3 minutes 5 Scale will turn off after 5 minutes 15 Scale will turn off after 15 minutes 3 0 Scale will turn off after 30 minutes F2 prt P mode Print Au on Data out pu...

Page 20: ...9600 Baud 8 data bits No Parity Connector 9 pin socket From C100 Pin 2 Output Pin 3 Input Pin 5 Signal Ground To LP50 Pin 2 Input Pin 3 Output Pin 5 Signal Ground Connection to the LP 50 printer Set P...

Page 21: ...will be show Ensure the platform is empty and wait for stable indicator Press to confirm Display will be show Set weight value will be required Enter the value by using numeric keys Press to confirm D...

Page 22: ...tor Press to confirm Display will be show Apply the weight on the display indicated in chart below wait for stable indication Then press to confirm Display will be show Apply the weight on the display...

Page 23: ...n Pi n Enter the password Default password is 0000 and press to confirm Pi n Select the scale by pressing which should be configured and press to confirm Tech local remote Use the key to select the we...

Page 24: ...cale division 2 5 10 20 50 Di v I nc 5 Set division I nc 10 I nc 20 I nc 50 A 2t Azn 0 5d Automatic zero tracking Azn 1d Azn 2d Azn 4d 0 Auto 0 auto 0 Zero setting range after switching on the scales...

Page 25: ...out of range Make sure platform empty and check the pan is installed proper Check the load cell connectors Err 9 Unstable Reading Check any air variation vibration RF noise and touching some where Che...

Page 26: ...C100 Series Technical Manual 23 11 DRAWING...

Page 27: ...0034 40 kg 14 Star Screw 4 M4x16 15 Load cell lower bracket 1 16 Self thread Screw 4 4x10 17 Insulated Washer 4 8x3 1x1 2t 18 Power Socket 1 19 Power Socket Spacer 1 20 Power Switch 1 899 300077 21 Fo...

Page 28: ...cells Connection 4 wire connection to load cells plus shield Maximum cable length 15 Feet Termination DB9 plug on scale Base Specification for Local Scale Load Cell Model No C2X1 Rated Capacity 6 50...

Page 29: ...NOTES...