4.3 Important Advice for Storage
CAUTION
• Before being loaded, the unit must have reached the set point temperature.
• The load should only be stored in the provided baskets and should not exceed the storage areas. The
total amount of stored blood bags should be evenly distributed on the 5 baskets (max. 20 blood bags
per basket) (image 6). Do not overload the unit.
• Different blood or plasma types should be sorted and stored in a manner for the User to identify these
with easy.
• The loading must never obstruct the air-inlet-openings.
• Do not store warm items in the devices. They are not designed for fast cooling of warm items.
• Avoid long and / or frequent lid openings to prevent that the inner temperature rises to high.
• Make sure that the lid is correctly closed by using the supplied latches and that the lid gasket seals cor-
rectly against the cabinet.
• The MB3000G is designed as storage unit for blood bags. It is not designed for cool down. The bags
should therefore be precooled at less than +6°C before storing them in the MB30000G.
• The maximum storage capacity of the MB3000G is 45 L of blood (100 blood bags of 500ml filled with
450ml each).
• Store the blood bags in upright position (image 5).
image 5: upright position of blood bags
image 6: max. storage capacity