DIAGNOSTICS
8
Display of the current status of the appliance
(only for service purposes).
q
PARAMETERS
8
Display of the current setting of all parameters
5.5 Status- and Warning messages
• The electronics generate various status or warning messages depending on the setting or the condition
of the appliance.
• If one or several messages are to be displayed at the same time in one line, these messages appear
alternately on the display.
w
HISTORY
This message indicates that a new entry has been made in the alarm
history list. The message disappears as soon as the history has been
displayed once with the function “SHOW HISTORY”.
w
ALARM TEST
Appears during an alarm test. The message disappears as soon as
the test has finished.
w
AMB ..°C
If an ambient temperature sensor is connected to the appliance and
the display in the EXTRA menu activated (
3
- EXTRA - AMBIENT
PROBE), this message appears followed by the currently measured
ambient temperature in the display.
w
REG ..°C
If the display of the regulating sensor is activated in the EXTRA
menu
(
3
- EXTRA - ADD. PROBE - REGULATION) this message appears
followed by the currently measured control temperature in the dis-
play.
w
SERVICE
If the appliance has been running for a preset time this message
appears on the display to indicate that a cleaning or maintenance
procedure is required. The message can be cancelled with the “HIDE
SERVICE WARNING” function in the SETTINGS menu.
w
HIGH AMBIENT TEMP.
If an ambient temperature sensor is connected to the appliance and
if this value rises above a given value this message appears on the
display. The appliance will however function normally. The maximum
value stored in the electronics depends on the temperature class of
the appliance in question and cannot be changed.
w
LOW AMBIENT TEMP.
analogous to “HIGH AMBIENT TEMP.” however with minimum value.
Summary of Contents for MB 3000 G
Page 1: ...820 9505 61 ed0821 Instructions for Use MB 3000 G ICE LINED BLOOD STORAGE REFRIGERATOR...
Page 2: ...2...
Page 3: ...3...
Page 6: ...6 1 Model Overview Group Model name Ice lined Blood storage Refrigerators MB3000G...
Page 31: ...31...
Page 32: ...B Medical Systemes S r l op der Hei 17 L 9809 Hosingen Luxembourg...