20
5 Electronics setup
5.1 Button functions
•
Pressing the MENU-button
3
gives you access to the menu area. Choose the respective menu using
the UP- and DOWN-keys. The selection is confirmed with the ENTER-button.
•
Press the BACK-key
)
to return to the previous screen or menu.
•
Navigation within the menus and changing of the settings are done using the UP-
p
, DOWN-
q
, LEFT-
t
und RIGHT
u
keys.
•
Use the ENTER-key
8
to confirm the actual selection. This applies both for menu selection and for para-
meter value. The change of a parameter only becomes effective when confirmed with the ENTER-key.
•
In the case of an alarm, the acoustic alarm can be switched off for a specified time with the “mute but-
ton”. In Menu-mode the “mute button” is out of function.
5.2 Setting the language
• The factory language setting of the electronics is ENGLISH. To convert to another language, proceed as
follows :
3
press the MENU-key.
q
select EXTRA
8
q
LANGUAGE
8
q
select your language
8
5.3 The “SETTINGS”-Menu
• The parameters contained in the “SETTINGS” menu partially intervene in the function of the appliance.
Access to the SETTINGS menu can therefore be protected by a password. It is recommended to use this
protection to limit access to the parameters.
3
press the MENU-key
w
select “SETTINGS”
8
w
USER PASSWORD :
0XXX
8
In the factory setting this password is set to the
value “0000” so that by pressing the
8
- button you
get directly into the SETTINGS menu.
q
SET POINT
8
w
SET : X.X °C
The temperature set point is the set temperature of
the refrigerator. The set point cannot be changed.
q
DATE/TIME
8
w
SET DATE/TIME
8
w
HH:MM
dd MMM yy
8
Setting current time and date.
t
SELECT 12h/24H
8
w
12h DISPLAY
pq
24h DISPLAY
8
Changing between 12-hour and 24-hour display
q
ALARM TEST
8
w
ALARM TEST ACTIVATION
8
w
ALARM TEST
8
This function is used for simulation a temperature
alarm situation. Pressing the ENTER-button will
launch the following alarm test cycle:
1. Phased increase of the internal temperature until
Summary of Contents for MB 3000 G
Page 1: ...820 9505 61 ed0821 Instructions for Use MB 3000 G ICE LINED BLOOD STORAGE REFRIGERATOR...
Page 2: ...2...
Page 3: ...3...
Page 6: ...6 1 Model Overview Group Model name Ice lined Blood storage Refrigerators MB3000G...
Page 31: ...31...
Page 32: ...B Medical Systemes S r l op der Hei 17 L 9809 Hosingen Luxembourg...