14
3.2 Final position
• The MB3000G is equipped with 4 bearing rolls.
Once it is in its final position, the feet must be
unscrewed to immobilise it and stop it rolling away
(image 3). Screw the feet with the provided key
until the rolls do not touch the floor anymore.
3.3 Initial cleaning and disinfection
CAUTION
• The use of hydrogen peroxide H2O2 (VHP) for the disinfection (sterilization) of the appliances is not suit-
able. This procedure may be applied exclusively with dedicated appliances.
• Other than the cleaing procedure described in this manual, follow the internal policy and procedures.
• Before using the refrigerator it should be cleaned inside and outside.
• For cleaning, use only gentle cleaning agents. Never use aggressive or caustic cleaning agents, scour-
ing powder, steel wool, abrasive sponges or chemical solvents. When cleaning, make sure that no fluids
of any kind run into the ventilation housing.
• For disinfecting, we recommend all the surface disinfecting agents commonly used by the customer, pro-
vided they are recommended by the national organizations. For disinfecting small areas, we recommend
using a concentrated alcoholic agent.
• Refer to section 8 for more information on cleaning.
3.4 Connection of remote alarm
• Connect the remote alarm located on the back of the unit to your surveillance system by respecting the
connecting diagram in section 7.2.
foot
bearing roll
image 3: handling of the feet
Summary of Contents for MB 3000 G
Page 1: ...820 9505 61 ed0821 Instructions for Use MB 3000 G ICE LINED BLOOD STORAGE REFRIGERATOR...
Page 2: ...2...
Page 3: ...3...
Page 6: ...6 1 Model Overview Group Model name Ice lined Blood storage Refrigerators MB3000G...
Page 31: ...31...
Page 32: ...B Medical Systemes S r l op der Hei 17 L 9809 Hosingen Luxembourg...