B medical systems MB 3000 G Instructions For Use Manual Download Page 17

17

4   Use and Operation 

WARNING

• The devices must only be used by adults. Do not allow children to play with it or touch the controls.
• Do not store flammable or explosive substances inside these appliances (e.g. aerosols with flammable

propellant). These may cause explosion or fire.

• Make sure that no sharp or pointed objects come into contact with the cooling system. The

cooling system contains flammable refrigerant. Both the appliance and the products stored
inside can be severely damaged if the system starts to leak.

• Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet. This may

cause electric shock or injury by accidental contact with moving parts.

• It is of utmost importance that the icelining elements are adequately prepared. Make sure that all these

elements are in place before starting the unit. 

• Make sure that the lid is securely closed by using the latches (refer to 4.2).
• Note that the unit may be locked using a key. Always ensure that the key is available to the user.

• It is of utmost importance to set up a password during the first start up procedure. Make sure that the

password is available to the person responsible.

CAUTION

• The sound intensity level of the unit is lower than 70dBA (measured at a distance of 1 meter)

4.1  Switching on

To switch the unit on, turn the key switch in horizontal position. The green power LED lights up.

After the self-test, the temperature inside the refrigerator is displayed.

The compressor will only switch on a certain time after the self-test. 

•  Make sure that during operation the drainage plug is closed properly.

4.2  Opening and closing

•  The unit comes with two latches, one of which contains a key lock system.
•  To open the unit, pull the clip up and remove the latch from the groove in the lid.
•  To close the unit, pull the clip up, hook the latch in the groove of the lid and pull the clip down. 
•  Once the unit has been securely closed, the user may lock the unit using a common key lock system.

image 4: latch and lock mechanism

Open latch

Closed latch and

key lock

Summary of Contents for MB 3000 G

Page 1: ...820 9505 61 ed0821 Instructions for Use MB 3000 G ICE LINED BLOOD STORAGE REFRIGERATOR...

Page 2: ...2...

Page 3: ...3...

Page 4: ...rtified according to the safety standard IEC 60335 1 This equipment is designed to be safe at least under the following condi tions based on the IEC 60335 1 Indoor use Altitude up to 2000 m Temperatur...

Page 5: ...Storage_______________________________________18 4 4 Control Panel ____________________________________________________19 4 4 1 Overview ________________________________________________19 4 4 2 Displa...

Page 6: ...6 1 Model Overview Group Model name Ice lined Blood storage Refrigerators MB3000G...

Page 7: ...ce lining consists of icepacks with water freezing during its operation During periods of power failure the ice bank acts as cold storage to protect the goods stored in the MB3000G The unit keeps the...

Page 8: ...requency other than that on the nameplate may cause fire or electric shock Do not store flammable or explosive substances inside these appliances e g aerosols with flammable propellant These may cause...

Page 9: ...he technical data see chapter 11 with a maximum relative humidity of 95 The device must be set up in a dry well ventilated place Avoid direct sunlight or locating it close to a heat source The locatio...

Page 10: ...ubmitting the delivery note or proof of purchase Do not operate a unit that has been damaged in transit If you are unsure contact your sales outlet and ask them The unit must be transported in an upri...

Page 11: ...e indoors in hospitals blood donation services blood transfusion centers and clinical laboratories The devices are designed for an ambient temperature range from 10 C to 43 C under operation and from...

Page 12: ...od or blood components The devices are not intended to be used for the cooling of foodstuffs or any other products not covered in the intended purpose The devices are not intended to be operated outsi...

Page 13: ...ectric shock may result due to corrosion The wall socket must be easily accessible When the unit is in its final position it is important to immobilise the unit using the feet CAUTION Optimal function...

Page 14: ...refrigerator it should be cleaned inside and outside For cleaning use only gentle cleaning agents Never use aggressive or caustic cleaning agents scour ing powder steel wool abrasive sponges or chemi...

Page 15: ...s may cause electric shock When disconnecting always pull the plug and never the cable Pulling the cord may result in electric shock or fire by short circuit CAUTION Before connecting and switching th...

Page 16: ...of the cabinet into the lining compartment between the inner and outer cabinet Watch out to introduce the holders in the correct way and not to damage the evapo rator tubes To ensure the best transpo...

Page 17: ...that the unit may be locked using a key Always ensure that the key is available to the user It is of utmost importance to set up a password during the first start up procedure Make sure that the passw...

Page 18: ...s Do not store warm items in the devices They are not designed for fast cooling of warm items Avoid long and or frequent lid openings to prevent that the inner temperature rises to high Make sure that...

Page 19: ...e disabled through the settings The temperature values are displayed with a resolution of 0 1 Refer to chapter 6 4 for more informa tion 4 5 Temperature Display During normal operation the display sho...

Page 20: ...MENU key q select EXTRA 8 q LANGUAGE 8 q select your language 8 5 3 The SETTINGS Menu The parameters contained in the SETTINGS menu partially intervene in the function of the appliance Access to the...

Page 21: ...nction q HIDE SERVICE WARNING 8 This function is for cancelling the service message q USER PASSWORD 8 Changing the password to access the SETTINGS MENU w CHANGE USER PASSWORD 0XXX XXXX 8 Enter the new...

Page 22: ...pressing the button during a power failure 100 approx 10 Sec 50 approx 5 Sec see chapter Alarm situations q VOLTAGE IND 8 w LINE VOLTAGE pq NONE 8 Shows the input voltage of the appliance on the dis...

Page 23: ...is connected to the appliance and the display in the EXTRA menu activated 3 EXTRA AMBIENT PROBE this message appears followed by the currently measured ambient temperature in the display w REG C If th...

Page 24: ...interrupted with the appliance switched on A power failure alarm is stored in the alarm history The alarm mes sage disappears as soon as the power supply is restored If one of the following error mess...

Page 25: ...the corresponding alarm message appears on the display 6 3 Alarm History The alarm history list contains all relevant data on temperature and powerfail events The list contains up to 20 alarm situatio...

Page 26: ...disposed of separately in the case of a fault 7 2 Remote contact There are two terminals with three floating contacts on the back of the unit that can be used to trigger an additional external alarm...

Page 27: ...de and outside For cleaning use only gentle cleaning agents Never use aggressive or caustic cleaning agents scour ing powder steel wool abrasive sponges or chemical solvents When cleaning make sure th...

Page 28: ...cabinet frame Close the lid Connect the power supply plug to the wall socket and switch on the MB3000G Allow it to run and observe the cabinet temperature Put back the contents into the MB3000G from...

Page 29: ...lly read and follow the information and advice below Problem Interior temperature too hot Please check if large quantities of warm goods have been placed inside the unit causing a surge of the interio...

Page 30: ...cold alarm C Preset warm alarm C Ambient operating temperature range C Storage and transport temperature not operational C Max Humididty range Altitude Refrigerant Type Refrigerant quantity g Hold ove...

Page 31: ...31...

Page 32: ...B Medical Systemes S r l op der Hei 17 L 9809 Hosingen Luxembourg...